Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Goerz SERVOGOR 790 Linearity (1yr) CAL VER /5520 DATE: 05-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 11 NUMBER OF LINES: 63 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # GOERZ SERVOGOR 790 Service Manual #331 0065 03 - 11/87 # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F V 1.002 ASK+ X 1.003 HEAD LINEARITY: [MEM2] Input 1.004 RSLT =LINEARITY: [MEM2] Input 1.005 M5520 0V AX 1.006 5520 0V S 2W 1.007 DISP Set [MEM2] Range to 1V/cm. 1.008 DISP Adjust the [MEM2] ZERO for 0% horizontal deflection. 1.009 MESS Slew the 5520A for 1cm scale deflection. 1.010 5520 1 1.00V 0.25/ 2W 2.001 MESS Slew the 5520A for 2cm scale deflection. 2.002 5520 1 2.00V 0.25/ 2W 3.001 MESS Slew the 5520A for 3cm scale deflection. 3.002 5520 1 3.00V 0.25/ 2W 4.001 MESS Slew the 5520A for 4cm scale deflection. 4.002 5520 1 4.00V 0.25/ 2W 5.001 MESS Slew the 5520A for 5cm scale deflection. 5.002 5520 1 5.00V 0.25/ 2W 6.001 MESS Slew the 5520A for 6cm scale deflection. 6.002 5520 1 6.00V 0.25/ 2W 7.001 MESS Slew the 5520A for 7cm scale deflection. 7.002 5520 1 7.00V 0.25/ 2W 8.001 MESS Slew the 5520A for 8cm scale deflection. 8.002 5520 1 8.00V 0.25/ 2W 9.001 MESS Slew the 5520A for 9cm scale deflection. 9.002 5520 1 9.00V 0.25/ 2W 10.001 MESS Slew the 5520A for 10cm scale deflection. 10.002 5520 1 10.00V 0.25/ 2W 11.001 MESS 11.002 M5520 * 11.003 END